SETTING
YOUR COMPUTER TO THE GDV CAMERA
After installing ZIPSHOT and GDV programs you should calibrate
the programs to the particular GDV Camera. For this:
-
Check the "BRIGHTNESS" setting in the Manual/Specifications
and adjust the recommended value in ARCSOFT ZIPSHOT Program.
-
Close ARCSOFT ZIPSHOT Program.
-
Open GDV Grabber Program.
-
Clear the GDV Lens with alcohol.
-
Place the GDV Test-object at the middle of the GDV Lens
inside the Cover.
-
Take 10 beograms of the test-object. They should look as
a dark ring with black ring inside. In case of distorted
ring or lightning inside check the position of the test-object
and clean the GDV Lens.
-
Save beograms in the GDV/New/Test folder.
-
Process beograms in the GDV Processor.
-
If there are "noise" around the test-object, click
Screen/Clear All and load the beograms again. Decrease the
Level Background from 250 to 240 and process beograms. If
there are "noise" again, repeat the procedure
and put 230, and so on till you see clear field around the
beogram. Usually we do not use Background Level less than
220.
-
Open all GDV Programs and put selected Background Level.
Close GDV Programs.
-
Open GDV Diagram Program. Press CALIBRATION (NORMALIZER)
button. Choose and select beograms of the test-object and
press OK. Wait for processing. This program calculates sector
parameters for the test-object beograms that work as calibration
references. In case if Coefficient of variation does not
exceed 4-5 press "Exit & Save". If Coefficient
of variation exceeds 5, press "Exit" and try to
get reproducible beograms of the test-object. If you fail
after a set of trials, repeat this procedure the next day
or contact Kirlionics Technologies International.
We
recommend calibrate the GDV Camera every 6 month of operation.